EN – TF03 – Handling
- If damage or an open seal is detected, sterilization may be compromised and the instrument should be discarded and replaced.
- Keep an appropriate inflation device with a pressure gauge available.
- Check the inflation device for appropriate pressure and volume.
- Keep enough filled liquid (e. g. sterile low viscosity contrast media; sterile, isotonic saline solution; or a mixture of both) available.
- To support applications with a guide wire, keep an appropriate guide wire available.
- Open packaging and carefully remove the pouch.
- Inspect the catheter to any damage. If you note any problems, exchange the balloon for a new one.
- Connect a syringe to the Luer-Lock and apply a moderate vacuum and remove protective tubing.
- To improve the slide properties inside the endoscope, spray the balloon with silicone. Only use silicone, other lubricants may damage the product.