EN – TF03 – Handling

  • If damage or an open seal is detected, sterilization may be compromised and the instrument should be discarded and replaced. 
  • Keep an appropriate inflation device with a pressure gauge available.
  • Check the inflation device for appropriate pressure and volume.
  • Keep enough filled liquid (e. g. sterile low viscosity contrast media; sterile, isotonic saline solution; or a mixture of both) available.
  • To support applications with a guide wire, keep an appropriate guide wire available.
  • Open packaging and carefully remove the pouch.
  • Inspect the catheter to any damage. If you note any problems, exchange the balloon for a new one. 
  • Connect a syringe to the Luer-Lock and apply a moderate vacuum and remove protective tubing.
  • To improve the slide properties inside the endoscope, spray the balloon with silicone. Only use silicone, other lubricants may damage the product.