Insertion into the Biopsy Channel
- Insert the device into the biopsy channel until it comes out of the distal part of the endoscope.
- Once under direct visualization, you are ready for the next step.
Deployment of the clip
- The clip is deployed closed. To open the clip, move the slider distally (away from the thumb ring)

Rotation of the clip
- To rotate the clip, in order to adjust the angle between the clip and the lesion, just rotate the complete handle in any direction.

Closing the clip
- To close the clip jaws on the desired location move the slider proximally (close to the thumb ring) until the first tactile resistance is felt in the handle.

- To permanently deploy the clip, continue moving the slider proximally (toward the thumb ring) beyond the tactile resistance point. At this point, click may be heard or felt.

Removal of the Delivery System
- Remove the whole delivery system from the biopsy channel and check the result with the endoscope, if needed repeat the whole procedure with other clips.